DTS-60A O
PERATORS
M
ANUAL
42
4.3 Standard Test Execution
If a standard test is selected for execution, the system will proceed with
the test initialisations and furthermore it will display the Electrode Spacing
warning message, Figure 4-7. The user must be sure for the current
Electrode Spacing; if not he must check and adjust it according to the
warning message indication. The only active keys are the ‘MENU’ key and
the ‘ENTER’ key.
Figure 4-7 – Electrode Spacing Warning Message
By pressing the ‘ENTER’ key the system proceeds to the selected Test
Program execution.
Figure 4-8 – Test Execution User Interface
The system screen for the Test execution, Figure 4-8, includes all the
information concerning the Test flow. More specifically, the 1
st
line indicates
the selected Test Name. The 2
nd
line indicates the Test Timer, which is used
as increasing timer (seconds resolution) during voltage rising, as well as a
decreasing timer for wait periods. The 3
rd
line indicates the current Stage of
execution, as well as the desired Voltage value and Duration. Obviously, the
TEST: IEC-156/95
PROGRAM BLOCKED...
DURATION:
00:00:00
Test Name
Timer Informations
Test Execution is Blocked!
Electrode Spacing Warning Message!
User Interface Help Message
Electrodes:
Space:
2.50±0.05mm
GO TO MENU
MENU
STARTUP WAIT TIMER...
DURATION:
0.0
00:04:39
Test Program Name
Count Down Timer
Test Messages
Current High Voltage Output
Stirrer Indicator
System Messages (Time/Date/Temperature. etc)
kV
STIR
TEST: IEC-156/95
12:10 01/08/05 65 °F
STAGE
1
Stage Indicator
Содержание DTS-60A
Страница 1: ...DTS 60A Dielectric Oil Test Set Operator s Manual Version 1 20 HighVoltage INC ...
Страница 3: ...DTS 60A OPERATORS MANUAL ...
Страница 5: ...DTS 60A OPERATORS MANUAL ...
Страница 7: ...DTS 60A OPERATORS MANUAL ...
Страница 10: ......
Страница 48: ...DTS 60A OPERATORS MANUAL 38 ...
Страница 56: ...DTS 60A OPERATORS MANUAL 46 Figure 4 11 Test Program Execution Example ...
Страница 57: ...DTS 60A OPERATORS MANUAL 47 ...
Страница 77: ...DTS 60A OPERATORS MANUAL 67 prg 0 stage 27 003 0 kV s breakdown 00 00 00 00 10 00 00 10 00 end store prg 0 end ...
Страница 81: ...DTS 60A OPERATORS MANUAL 71 ...
Страница 87: ...DTS 60A OPERATORS MANUAL 77 ...
Страница 106: ...DTS 60A OPERATORS MANUAL 96 Figure 8 17 Reading specific result id Figure 8 18 history erase Command ...
Страница 111: ...DTS 60A OPERATORS MANUAL 101 ...