![High Voltage DTS-60A Скачать руководство пользователя страница 42](http://html.mh-extra.com/html/high-voltage/dts-60a/dts-60a_operators-manual_2130381042.webp)
DTS-60A O
PERATORS
M
ANUAL
32
3.4 Simple Test Execution
If the ‘ENTER’ key is selected, the system executes the Simple
Test. If the user presses any key, the execution will be immediately
interrupted and the Simple Test will be cancelled. Otherwise, the Simple
Test will continue the voltage increasing until the detection of a
breakdown, Figure 3-6. Then the system will be halted.
The user should press any key in order to continue. The system
will ask the user if a new test for the same insulator oil sample is
desired, Figure 3-6. If the user presses the ‘ENTER’ key a new test
process will be initiated, Figure 3-2.
Figure 3-6 – Proceed to a new test
1.0
kV
STIR
STAGE
1
HN on
12:10 01/08/05 65 °F
48.2
PRESS ANY KEY...
kV
STIR
STAGE
1
High Voltage ON indicator
Breakdown Indicator
TEST:SIMPLE TEST
TIMER:
00:00:02
VOLTAGE:
0.5 kV/s
TEST:SIMPLE TEST
TIMER:
00:01:36
VOLTAGE:
0.5 kV/s
Содержание DTS-60A
Страница 1: ...DTS 60A Dielectric Oil Test Set Operator s Manual Version 1 20 HighVoltage INC ...
Страница 3: ...DTS 60A OPERATORS MANUAL ...
Страница 5: ...DTS 60A OPERATORS MANUAL ...
Страница 7: ...DTS 60A OPERATORS MANUAL ...
Страница 10: ......
Страница 48: ...DTS 60A OPERATORS MANUAL 38 ...
Страница 56: ...DTS 60A OPERATORS MANUAL 46 Figure 4 11 Test Program Execution Example ...
Страница 57: ...DTS 60A OPERATORS MANUAL 47 ...
Страница 77: ...DTS 60A OPERATORS MANUAL 67 prg 0 stage 27 003 0 kV s breakdown 00 00 00 00 10 00 00 10 00 end store prg 0 end ...
Страница 81: ...DTS 60A OPERATORS MANUAL 71 ...
Страница 87: ...DTS 60A OPERATORS MANUAL 77 ...
Страница 106: ...DTS 60A OPERATORS MANUAL 96 Figure 8 17 Reading specific result id Figure 8 18 history erase Command ...
Страница 111: ...DTS 60A OPERATORS MANUAL 101 ...