4.4 Standard Test Example
A detailed example of a Test Execution is illustrated in Figure 4-14.
Figure 4-14 –
Test Program Execution Example
42
STARTUP WAIT TIMER...
0.0
Test Program Name
Count Down Timer
Test Messages
Current High Voltage Output
Stirrer Indicator
System Messages (Time/Date/Temperature. etc)
kV
STIR
STAGE
1
Stage Indicator
AFTER BD WAIT TIMER...
42.2
Test Program Name
Count up Timer for current stage duration
Rate of rise from 0kV to breakdown
System Messages
kV
Current Stage Indicator
STAGE
1
Blinking Breakdown Indicator
STIR
Question Message
Test Program Name
Test is finished!
System Messages
Test Name
Timer Information
Test Ready to start.
Electrode Spacing Warning Message!
User Interface Help Message
1.0kV/s > BREAKDOWN
30.7
Test Program Name
Count up Timer for current stage duration
Rate of rise from 0kV to breakdown
System Messages
kV
STAGE
1
HV on
Current Stage/ High Voltage ON Indicator
TEST:ASTM-D1816/84
TIMER:00:00:00
TEST READY...
ENTER > PROCEED TEST
Electrodes:
Spacing:0.04in
TEST:ASTM-D1816/84
TIMER:00:00:17
12:10 01/08/16 65°F
12:10 01/08/16 65°F
12:10 01/08/16 65°F
TEST:ASTM-D1816/84
TEST:ASTM-D1816/84
TIMER:00:00:30
TIMER:00:00:21
TEST RESULTS
TEST:ASTM-D1816/84
NUM. BD: 5
MEAN: 22.92kV
STD: 00.28
PRESS ANY KEY...
Test Results
?
12:10 01/08/16 65°F
Print
Results?
TEST RESULTS
NUM. BD: 5
YES
TEST:ASTM-D1816/84