3.3 Simple Test Controls
After adjusting the electrodes and closing the lid, the system will still be halted until the user presses
the
‘ENTER’
key.
Then the system proceeds to the Simple Test Main Screen in which the user is expected to set the
voltage ramp for this next test stage.
This screen displays the following information (as shown in Figure 3-4):
1
st
line: Test Name
2
nd
line: Adjustable voltage ramp (blinking)
3
rd
line: Stage Count for this test run
Current high voltage output (large digits in kV)
Stirrer icon (if enabled)
Last line: Current time, date and temperature or test messages
0.0
kV
STIR
STAGE
1
Program Name is Simple Test
Voltage Rate of Rise Setup (Blinking)
Current High Voltage Output
Stirrer Indicator
System Messages
(Time/Date/Temperature. etc)
Stage Indicator
TEST:SIMPLE TEST
TIMER:00:00:00
VOLTAGE:
0.5 kV/s
12:10 01/08/16 65°F
Figure 3-4 –
Simple Test Main Screen
The following controls are active:
1. Use the
‘UP’
and
‘DOWN’
keys to change the voltage rate of rise setup for the next test
stage
2. Use the
‘BACK’
key for enabling/ disabling the Stirrer in between test stages
3. Use the
‘ENTER’
key for choosing the voltage ramp and starting the Simple Test
4. Use the
‘MENU’
key to end the Simple Test and proceed to the Main Menu.
28