![Hico Medical Systems HICO-VARIOTHERM 550 Скачать руководство пользователя страница 63](http://html1.mh-extra.com/html/hico-medical-systems/hico-variotherm-550/hico-variotherm-550_instructions-for-use-manual_2129330063.webp)
Technical data and accessories
HICO-VARIOTHERM 550
63
11 Technical data and accessories
11.1 Unit
HICO-VARIOTHERM 550
Article No. (REF):
540001 (220 V/230 V) / 540021
(115 V)
Rated voltage:
230 VAC 50/60 Hz / 115 VAC
50/60 Hz / 220 VAC 60 Hz
Power input:
320 W
Current consumption:
approx. 1.5 A (220 V/230 V)/ 3 A
(115 V)
Nominal value range:
15-39 °C
Safety shutdown:
41.1-41.5 °C
(autonomous safety-
shutdown) This results in a max. surface
temperature on the pad < 41 °C
Correction value:
0.5 °C
(Water temp. -temp.-display)
Sensor element:
2 x NTC 5 K
Pump capacity:
max. 5.5 l/min., max. 0.21 bar
Warming up time:
approx. 5-10 min.
(20-37 °C)
Cooling down time:
approx. 5-10 min.
(20-15 °C)
Fuse rating:
2x T 3,15 A; L 250 V (220 V/230 V)
2x T 5 A; L 250 V (115 V)
Class/type of protection:
I, BF
(defibrillation protected)
Type of protection IP:
IP X1
(drip proof)
Power supply:
Must correspond with VDE 0100 Part
710:2012-10 "
Electrical safety in medical
environments"
Risk class (93/42/EEC):
II b
Ambient temperature:
10-30 °C (
hyperthermia operation
)
10-23 °C
(hypothermia operation)
Relative air humidity:
approx. 30-70 %
Transport/storage temperature:
3-60 °C
Air pressure:
700-1060 hPa
Tank volume:
approx. 0.5/0.8 l
(MIN/MAX)
Permissible height difference:
max. 1 m
(Unit/Water pad)
Dimensions WxHxD:
approx. 200 x 290 x 440 mm
Weight:
approx. 17 kg (empty)
Noise emission:
approx. 50 dB(A) (1 m)
Alarm level:
> 55 dB(A) (3 m)
Test basis:
Medical product directive 93/42/EEC,
DIN EN 60601-1, DIN EN 60601-1-2, DIN
EN 80601-2-35
UMDNS code:
12-075
Subject to technical changes