Control and operation
HICO-VARIOTHERM 550
49
6.3.2.5 Display message: CHECK UNIT
CUSTOMER SERVICE
The unit triggers this alarm in case of various defects. The
display shows "CHECK UNIT" and "
CUSTOMER SERVICE",
the yellow fault lamp (13) flashes and a pulse signal tone sounds.
The audible alarm
cannot
be interrupted using the "Audio
paused" key (18).
•
Switch off the unit with the mains switch (16).
Take the unit out of service and have it inspected by a Service
engineer (Customer service / Medical engineering).
6.3.2.6 Mains failure alarm
The unit triggers this alarm if the power supply fails during
operation. The yellow fault lamp (13) lights up and a permanent
signal sounds. All other displays are without function. The power
accumulator in the unit receives the alarm for at least 10 minutes
without mains supply.
The alarm
cannot
be interrupted with the "Audio paused" key
(18).
•
Switch off the unit with the mains switch (16).
The alarm automatically goes out when the power supply returns.
The temperature set last is stored in the unit.
With restoration of the power supply, the temperature stored
must be controlled by the user and approved before the device
returns to normal operation (see chapter 6.2 Operation).