Commissioning
HICO-VARIOTHERM 550
21
4.3.3 Electrical connection
Danger caused by electrical current
Defective cables and/or plugs as well as faults in the power
supply can cause life threatening electric shock!
► Check the condition of unit cable and plug before connecting!
► To avoid the risk of electric shock, this unit must only be
connected to the mains with a protective earth conductor!
Please observe the following notes when connecting the unit
electrically to ensure safe and fault free operation:
•
Before connecting the unit, compare the electric data
(voltage
and frequency) on the type plate with the data of your mains
power supply. This data must match to prevent the unit from
being damaged
Please ask your electrics expert for advice, if in doubt.
•
The electric socket must be protected by a 16 A circuit breaker.
•
Use the cable supplied with the unit to connect the unit to the
mains supply. The unit socket is located on the rear side of the
unit (see chapter 5.1, Views of the unit, control and display
elements).
•
Power supply at the installation site must meet the
requirements for electrical systems in hospitals and medical
environments (see VDE 0100 Part 710:2012-10
"Low-voltage
electrical installations Part 7-710: Requirements for special
installations or locations - Medical locations.").