Hettich AG, Seestrasse 204a, 8806 Baech, Switzerland
Operator_Manual_Rotolavit II_und_Rotolavit II-S_en_rev2.5
Seite 50
33
Lid
detection
failure
Lid sensor has detected an opening of the
lid, but the lock sensor still signalizes the
lock is closed
Wrong lid detection of the lid
sensor
Wrong lock detection of the
lock sensor
System error
40
Program
reading
error
It was not possible to read the complete
program.
Program file is corrupted
Not enough dynamic memory
was available
41
Image
loading failed
Not all images could be loaded
An image is missing on the
flash
An image on the flash is
corrupted
42
EEPROM-
error
Loading data from EEPROM failed. (reading
not possible or checksum for the data is
incorrect)
EEPROM not initialized (login of
a service user needed)
Communication failure
Miscellaneous
50
Unknown
An error has occurred, but the kind of the
error could not be identified
Unexpected behavior
51
Program
interrupted
A running program was interrupted
Power interrupted during a
running program.
52
Program
aborted by
user
The program was aborted by the user
User has aborted the program
53
Imbalance
The program was stopped because of an
imbalance of the rotor
Rotor was not loaded
symmetrically
Positioning of the imbalance
sensor not correct
„Screen freeze:
If a screen freeze occurs, not the normal standby mode (push anywhere onto the dark display), perform
a main reset
Perform a MAINS RESET:
Switch off the mains switch (position "0", fig. 7.1, pos. 3).
Wait at least 10 seconds and then switch on the mains switch again to position "1".
Check the history for the last run and report the error code to your local service support
Before you open the lid with the release pin (cha. 4.3) you must check through the window into the lid
that the rotor is at a standstill position
If it is not possible to close the lid => check if there is a small part that has fallen into the lid lock
access hole (fig. 7.5), if so then contact your local service support