![Hettich ROTOFIX 46 Скачать руководство пользователя страница 40](http://html1.mh-extra.com/html/hettich/rotofix-46/rotofix-46_operating-instructions-manual_2127810040.webp)
EN
40/87
22
Centrifugation of materials or mixtures of materials with a density higher than 1.2 kg/dm
3
When centrifuging with maxim revolutions per minute the density of the materials or the material mixtures may not
exceed 1.2 kg/dm
3
.
The speed must be reduced for materials or mixtures of materials with a higher density.
The permissible speed can be calculated using the following formula:
[RPM]
speed
maximum
x
[kg/dm³]
density
Greater
1.2
=
)
(n
speed
Reduced
red
e.g.: maximum speed RPM 4000, density 1.6 kg/dm
3
RPM
3464
=
RPM
4000
x
kg/dm³
1.6
kg/dm³
1.2
=
n
red
In the exceptional case that the maximum loading indicated on the hanger is exceeded, the speed must also be
reduced.
The permissible speed can be calculated using the following formula:
[RPM]
speed
maximum
x
[g]
load
actual
[g]
load
maximum
=
)
(n
speed
Reduced
red
e.g.: maximum speed RPM 4000, maximum load 300 g, actual load 350 g
RPM
3703
=
RPM
4000
x
g
350
g
300
=
n
red
If in doubt you should obtain clarification from the manufacturer.
23 Rotor
recognition
Rotor recognition is carried out after each start of the centrifugation run.
If the rotor has been changed, the centrifugation run is canceled after rotor recognition. The rotor code (R xx) and the
maximum speed (n-max=xxxxx) of the rotor are displayed.
A further operation of the centrifuge is only possible after a single opening of the lid.
If the maximum speed of the rotor being used is less than the set speed, the speed is limited to the rotor's
maximum speed.
24 Emergency
release
The lid cannot be opened during power failure. An emergency release has to be executed by hand.
For emergency release disconnect the centrifuge from the mains.
Open the lid only during rotor standstill.
Only the plastic release pin provided may be used for emergency release.
See figure on page 2.
Switch off the mains switch (switch position "0").
Look through the window in the lid to be sure that the rotor has come to a standstill.
Insert the release pin horizontally into the hole (fig. 1, A). Push the release pin as far as possible, until the handle
can be turned to the left when pressing down the pin.
Open the lid.