14 Entering
centrifugation
parameter
If no key is pressed for 8 seconds long after the selection or during the input of parameters, the previous
values will be shown in the display. The input of parameter then has to be executed again.
Select the RPM or RCF display with the key
RCF
. RCF values are displayed in
> <
.
Select the desired parameters using the button
SELECT
and set using the knob .
In order to set continuous operation, the parameters
t/min
and
t/sec
must be set to zero with the knob.
Continual running is represented in the display by the following symbol, "
".
After input of all parameters, press the key
START / IMPULS
in order to store the adjustments on the programme
position
#
. As confirmation,
ok
will be displayed for a short period.
The data on the programme position # will be overwritten with every input of parameters and pressing of the
key
START / IMPULS
.
15 Programming
15.1 Programme
input/alteration
If no key is pressed for 8 seconds long after the selection or during the input of parameters, the previous
values will be shown in the display. The input of parameter then has to be executed again.
Select the RPM or RCF display with the key
RCF
. RCF values are displayed in
> <
.
Select the desired parameters using the button
SELECT
and set using the knob .
In order to set continuous operation, the parameters
t/min
and
t/sec
must be set to zero with the knob.
Continual running is represented in the display by the following symbol, "
".
The parameter
PROG STO
can be selected using the button
SELECT
and the desired program position set using
the knob .
Press the button
START / IMPULS
in order to store the setting on the desired program position.
ok
is
displayed briefly as confirmation.
If the key
START / IMPULS
is pressed without the parameter
PROG STO
being activated, the settings are always
stored in the program place
#
.
The previous data in the program position is overwritten during saving.
15.2 Programme
recall
Select the parameter
PROG RCL
using the button
SELECT
and set the desired program position using the knob
.
Press the button
START / IMPULS
. The centrifugation data of the selected program position is displayed.
The parameters can be checked by pressing the button
SELECT
.
To leave the parameter display press the button
STOP
or press no button for a period of 8 seconds.
16 Centrifugation
When the centrifuge is running, according to EN / IEC 61010-2-020, no persons, dangerous substances or
objects may be within the safety margin of 300 mm around the centrifuge.
If the permissible weight difference within the rotor loading has been exceeded, the drive shuts down during
the start-up, the unbalance display lights up, and
IMBALANCE
is displayed.
A centrifugation run can be stopped at any time by pushing the key
STOP
.
All parameters can be selected and altered during the centrifugation run (see Chapter “Entering centrifugation
parameter”).
You can switch-over at any time between the RPM and RCF display with the key
RCF
. The input of the
centrifugation radius is necessary if you are working with the RCF display.
If
OPEN OEFFNEN
is displayed, a further operation of the centrifuge is only possible after opening the lid
once.
If
R xx n-max xxxxx
is displayed, then no centrifugation run has taken place as the rotor was changed
beforehand, refer to Chapter " Rotor Identification ".
Turn on the mains switch. Switch position
.
Load the rotor and close the centrifuge lid.