Troubleshooting
43
Troubleshooting
Self-help measures
Fault report
Behaviour of the centri-
fuge
Possible causes and remedies
Setting posi-
tion flashing
after pressing
the "set" key
Setting with "+/-" keys
not possible, lid can be
opened, start and stop
O.K.
Key switch position prevents programming.
Move the key switch from position C to A or B, if there is no reaction,
request service.
Display re-
mains dark
The drive stops.
The rotor is slowing
down without decelera-
tion.
The lid does not open
The power supply is interrupted.
Is the power switch turned on?
Check the voltage connection.
If the voltage is O.K., inform your nearest customer service.
Displays break
down for short
periods
The drive suddenly stops.
The rotor is slowing down
with deceleration.
The power supply was interrupted for a short time.
1.
Turn off the power switch.
2.
Check that the power plug is correctly in the socket.
3.
Start the centrifuge again.
Unusually loud
running noises
Press the quick stop symbol key.
When the appliance is at a standstill, Iook for obvious causes and if
possible, eliminate. In other cases, request service.
Lid does not
open
Pressing on “lid open“
key to no effect.
Lid is not engaged properly or lid is distorted.
1.
Check that the appliance is connected to the power supply and
that it is switched on (display’s lights are on).
2.
If this has no effect, you can open the lid using the mechanical
emergency unlocking (see page 6).
–
Loud running noises.
1.
Stop the appliance. To do this, press the "quick stop" key, or in an
emergency, unplug from the power supply.
2.
Wait until the centrifuge is at a standstill.
3.
Check if the rotor is correctly loaded.
4.
Check to see if a broken vessel, a damaged rotor or a damaged
motor is causing the noises.
If you cannot recognise or eliminate the fault yourself, inform the cus-
tomer service.
"imbalance"
LED is alight,
intermittent
tone
The rotor is slowing
down with deceleration.
Imbalance switch triggered off.
1.
Open the appliance by pressing the "lid open“ key.
2.
Check if the rotor is correctly loaded .
3.
Check to see if a broken vessel or a damaged rotor has triggered off
the imbalance switch.
Display
showing
"OPEN" al-
though the lid
is closed.
No start-up possible.
The lid is not closed correctly.
Open the lid and repeat the locking process.
Содержание CRYOFUGE 6000i
Страница 1: ...CRYOFUGE 6000i CRYOFUGE 8500i Operating Instructions ...
Страница 4: ......
Страница 8: ...Safety Measures and Hazards 4 ...
Страница 12: ...Description of Appliance 8 ...
Страница 14: ...Instructions for Rapid Start up 10 ...
Страница 22: ...Transport and Installation 18 ...
Страница 42: ...Instructions for Temperature Control 38 ...
Страница 46: ...Maintenance and Service 42 ...