Instructions for Rapid Start-up
9
Instructions for Rapid Start-up
Switching on the centrifuge
1.
Switch on the machine’s power switch on the front right and wait for the appliance identification display.
2.
Wait for the yellow LED to the right of the lid symbol key to light up, press the key. After the lid unlocks,
lift it by hand.
3.
Check that the rotor is securely fixed and can be rotated fully!
4.
All retaining bolts must be lubricated. Operation is only permitted when all 6 buckets are inserted and are
of the same type!
5.
The buckets must be symmetrically loaded!
6.
Firmly close the lid but do not let it fall. The "Open" display disappears when the lid is correctly closed.
Programming the first operation
7.
Entering the order number of the inserted buckets:
Press the selection key with the hand symbol of the 4 function field until the “rotor” LED is alight and then
press the "set" key for all setting processes.
The last bucket order number entered will flash in the display and can now be changed by using the ar-
row keys (up, down). When the flashing and the inserted bucket numbers (engraved on all buckets) are
the same, the setting can be completed by pressing the "enter" key, or other parameters can be entered
one after the other (see 8. to 11.).
WARNING!
With the swing rotor, the order number of the buckets currently inserted must always be pro-
grammed, otherwise this could lead to the destruction of the appliance due to incorrectly entered
speeds.
8.
Setting the speed:
Press the hand symbol key of the 4 function field until the "speed" LED is alight. The speed position last
entered flashes and can be changed with the white direction arrows. The numerical value (rpm) can be
entered using the arrow symbol keys (to enter the RZB value, see "SEPACONTROL programming“).
9.
Pre-setting the running time:
Press the "hand" key of the "time/delay“ field ("time" LED must light up), set as above using direction ar-
row and "up" or "down" keys.
10.
Setting the temperature:
Press the "hand" key of the "temp/
D
T“ field ("temp" LED must light up), set as above under direction ar-
row and "up" or "down" keys.
11.
Acceleration and deceleration curves can be set in a similar way in the "accel/decel" field or can be de-
fined by the user.
12.
Saving the values entered and starting up the centrifuge:
Press the double function key "start/stop" within 2 minutes. After the programmed time has elapsed,
centrifugation ends automatically. Press the "start/stop" or the quick stop symbol key to stop prema-
turely.
Preparing for the next run
13.
Wait for release for the lid key (yellow LED is alight; rotor standing still). Press the lid symbol key, open
the lid and check the state of the vessels before removing them.
14.
Check vessels for damage, check the rotor and inserts for leaking fluids. If necessary, clean or disinfect
(see also under "Service and Maintenance“, page 39).
Further information can be found in the corresponding chapters!
Содержание CRYOFUGE 6000i
Страница 1: ...CRYOFUGE 6000i CRYOFUGE 8500i Operating Instructions ...
Страница 4: ......
Страница 8: ...Safety Measures and Hazards 4 ...
Страница 12: ...Description of Appliance 8 ...
Страница 14: ...Instructions for Rapid Start up 10 ...
Страница 22: ...Transport and Installation 18 ...
Страница 42: ...Instructions for Temperature Control 38 ...
Страница 46: ...Maintenance and Service 42 ...