![Heraeus CRYOFUGE 6000i Скачать руководство пользователя страница 18](http://html1.mh-extra.com/html/heraeus/cryofuge-6000i/cryofuge-6000i_operating-instructions-manual_2125625018.webp)
Transport and Installation
14
Screw in the stud bolts and safety hooks up to the top, so that they
do not get caught when the centrifuge is being rolled down.
Illustration 4:
Removing the stand feet
When rolling the centrifuge (observe weight) down from the pal-
ette, at least 2 adults are required (see Illustration 5). In case of
doubt, use a third person.
Once again ensure that the rollers run smoothly over the rails and
check the steering rollers are facing the rear.
Illustration 5:
Centrifuge being slowly rolled from the palette by
2 people
After installation at the location, re-attach the stand feet.
Observe requirements of the place in which the centrifuge is to be
installed!
Screw out the stud bolts, put the appliance feet underneath in the
centre and align the centrifuge so that all feet are equally burdened
(carry out the toppling test) and the transport rollers are approxi-
mately a hand’s width from the floor.
Illustration 6:
Re-attaching the stand feet
The final assembly of the safety hooks takes place after unpacking and installing the rotor and subsequent
levelling of the drive. (see “Installing the rotor“).
Illustration 4
Illustration 5
Illustration 6
Содержание CRYOFUGE 6000i
Страница 1: ...CRYOFUGE 6000i CRYOFUGE 8500i Operating Instructions ...
Страница 4: ......
Страница 8: ...Safety Measures and Hazards 4 ...
Страница 12: ...Description of Appliance 8 ...
Страница 14: ...Instructions for Rapid Start up 10 ...
Страница 22: ...Transport and Installation 18 ...
Страница 42: ...Instructions for Temperature Control 38 ...
Страница 46: ...Maintenance and Service 42 ...