24
Pulse Rate
Measurement range: 30~250 bpm
Resolution: 1 bpm
Accuracy: ±2 bpm or 2% (The larger are applied)
Perfusion Index
Display range: 0.1%~20%
Accuracy: 0.1%~1.0%, ±0.2%; 1.1%~20%, ±20%
Probe LED Specifi cations
Equipment data update period
As shown in the following fi gure. Data update period of slower average is 8s.
Operation Environment
Operating temperature: 0°C ~ 40°C
Relative humidity: ≤80%, no condensation
Atmosphere pressure: 86 kPa ~ 106 kPa
Transport and Storage Environment
Storage Temperature: -20°C ~ 70°C
Relative humidity: ≤93%, no condensation
Atmosphere pressure: 50 ~ 106 kPa
Power Requirement
Power supply: Three AA alkaline or Ni-MH batteries, or power adapter
Working time: work for 10 hours continuously at most
Wavelength
Radiant Power
RED
660±3nm
3.2mW
IR
905±10nm
2.4mW