19
• Do not pour liquid onto the equipment or accessories.
• Do not allow liquid to enter the case.
• Never use abrasive materials (such as steel wool or silver polish), or erosive cleaners (such as acetone
or acetone-based cleaners).
Caution:
If you spill liquid on the equipment or accessories, contact us or your service personnel.
Note:
To clean or disinfect reusable accessories, refer to the instructions delivered with the accessories.
8.2 Safety Checks
Before every use, or after your pulse oximeter has been used for 6 to 12 months, or whenever your pulse
oximeter is repaired or upgraded, a thorough inspection should be performed by qualified service personnel
to ensure the reliability. Follow these guidelines when inspecting the equipment:
• Make sure that the environment and power supply meet the requirements.
• Inspect the equipment and its accessories for mechanical damage.
• Make sure that only specified accessories are applied.
• Inspect if the alarm system functions correctly.
• Make sure that the batteries meet the performance requirements.
• Make sure that the pulse oximeter is in good working condition.
In case of any damage or abnormity, do not use the pulse oximeter. Contact your hospital’s biomedical
engineers or your service personnel immediately.
Cleaning
Please use medical alcohol to clean the silicone touching the finger inside of SpO2 probe with a soft cloth
dampened with 70% isopropyl alcohol. Also clean the being tested finger using alcohol before and after
each test. To clean your equipment, follow these rules:
1. Shut down the pulse oximeter and take the batteries out of the battery wharf.
2. Clean the display screen using a soft, clean cloth dampened with a glass cleaner.
3. Clean the exterior surface of the equipment using a soft cloth dampened with the cleaner.
4. Wipe off all the cleaning solution with a dry cloth after cleaning if necessary.
5. Dry your equipment in a ventilated, cool place.
Disinfecting
The applied parts touching the patients’ body are required to be disinfected once after each use. The recom-
mended disinfectants include: ethanol 70%, isopropanol 70%, glutaraldehydetype 2% liquid disinfectants.
Disinfection may cause damage to the equipment and is therefore not recommended for this pulse oxi-
meter unless otherwise indicated in your hospital’s servicing schedule. Clean the pulse oximeter before
disinfecting it.
Caution:
Never use EtO or formaldehyde for disinfection.
"dudu" voice should be heard.