5
360113-1
Troubleshooting
Error
message
number
Possible cause
Action
Continued
04,
06 to 09
A part in the lid lock circuit is faulty.
1
Clear the error message: Open the lid, then
turn off the power. After about 10 seconds, turn
the power back on.
2
Try the following tasks in order, testing after
each to determine if it solved the problem:
a
Replace the ribbon cable between the
control/display board and the power
supply board.
b
Replace the power supply board.
c
Replace the control panel.
21 to 26
The control panel is faulty.
1
Clear the error message: Open the lid, then
turn off the power. After about 10 seconds, turn
the power back on.
2
Replace the control panel.
FU/CCI - ERROR XX
When a message of this type appears, there is a problem related to the frequency converter.
Error
message
number
Possible cause
Action
60
A part in the frequency converter
circuit is faulty, resulting in a false lid
lock release signal.
►
Try the following tasks in order, testing after
each to determine if it solved the problem:
1
Replace the ribbon cable between the
control/display board and the power
supply board.
2
Replace the ribbon cable between the
frequency converter and the power supply
board.
3
Replace the frequency converter. For
instructions, see
Section 4.7, “Replacing
the frequency converter.”
4
Replace the power supply board.
5
Replace the control panel.
61
A part in the frequency converter
circuit is faulty, resulting in a
processing error.
1
Clear the error message: Open the lid, then
turn off the power. After about 1 minute, turn
the power back on.
2
Try the following tasks in order, testing after
each to determine if it solved the problem:
a
Replace the ribbon cable between the
control/display board and the power
supply board.
b
Replace the ribbon cable between the
frequency converter and the power supply
board.
c
Replace the frequency converter. For
instructions, see
Section 4.7, “Replacing
the frequency converter.”
d
Replace the power supply board.
e
Replace the control panel.