Troubleshooting
N < MIN XX
When this message appears, the rotor is rotating too slowly.
Error
message
number
Possible cause
Action
13
A connection in the motor circuit is
loose.
1
Clear the error message: Open the lid, then
turn off the power. After about 10 seconds, turn
the power back on.
2
Check the connection between the motor and
the frequency converter.
A part in the motor circuit is faulty.
1
Clear the error message: Open the lid, then
turn off the power. After about 10 seconds, turn
the power back on.
2
Try the following tasks in order, testing after
each to determine if it solved the problem:
a
Test the motor windings. If they are faulty,
replace the motor.
b
Replace the frequency converter. For
instructions, see
Section 4.7, “Replacing
the frequency converter.”
POWER INTERRUPT
When this message appears, the AC power was interrupted during operation, or was sensed as being
interrupted.
Error
message
number
Possible cause
Action
None
A connection between the control/
display board and the power supply
board is loose.
1
Clear the error message: Open the lid then
press the START WASH button.
2
Check the connection between the control/
display board and the power supply board.
A part in the control panel and power
supply circuit is faulty.
1
Clear the error message: Open the lid then
press the START WASH button.
2
Try the following tasks in order, testing after
each to determine if it solved the problem:
a
Replace the power supply board.
b
Replace the ribbon cable between the
control/display board and the power
supply board.
c
Replace the control panel.