10 Heating
with
electronic contact thermometer
Hei-Con
or contact thermometer
Connect the contact thermometer as described in "Set-up". Set the
Hei-Con
or contact
thermometer to the desired temperature. Now set the hotplate temperature as described in
chapter 9, and activate the hotplate function.
Note on heating time:
Higher hotplate temperatures will result in a shorter heat-up period.
Large differences between the hotplate temperature setting and the
selected nominal temperature of the
Hei-Con
temperature controller or
contact thermometer may result in temperature fluctuations. If such
fluctuation is not desirable, set the hotplate to a lower temperature.
This will lead to a longer heat-up period.
CLEANING AND MAINTENANCE
Use a cloth dampened with mild soapy water to clean the housing and the surface of the
instrument.
Note
Do not use chlorine bleach or other chlorine-based cleaning products,
abrasive cleanser, ammonia, steel wool or cleaning products with
metallic components under any circumstances. These will damage the
finish of the instrument.
The instrument is maintenance-free. Should repairs become necessary, please contact
an authorised Heidolph Instruments repair technician. Please contact your Heidolph
Instruments dealer or a Heidolph Instruments representation for more information
(see page 34).
Caution
Unplug the AC power cord before opening the housing.