Operation
25
Error message
Description
ERR_2
Incorrect entry (e.g. data buffer overflow)
ERR_3
Incorrect command
ERR_5
Syntax error in value
ERR_6
Impermissible value
ERR_32
The upper temperature limit is lower or the same as the
lower temperature limit.
▪
The RS 232 interface can be operated with or without a hardware
handshake (RTS/ CTS).
▪
Do not send commands in a package,
pause of at least 0.1 seconds between commands
▪
End every command with CR (Carriage Return, Hex=0D) or
LF (Linefeed, Hex=0A)
▪
The response from the chiller always ends with CRLF
▪
Date format XX.X or -XX.X (X stands for a number)
▪
As well as ”_“, ” “ (blank space) is also permissible
Cable and interface test for RS232
Computer
Thermostat
Signal
9-Pole Sub-D socket
25-Pole Sub-D socket
9-Pole Sub-D socket
Signal
with a
hardware
handshake
without a
hardware
handshake
with a
hardware
handshake
without a
hardware
handshake
with a
hardware
handshake
without a
hardware
handshake
RxD
2
2
3
3
2
2
TxD
TxD
3
3
2
2
3
3
RxD
DTR
4
-
20
-
4
-
DSR
Signal
Ground
5
5
7
7
5
5
Signal
Ground
DSR
6
-
6
-
6
-
DTR
RTS
7
-
4
-
7
-
CTS
CTS
8
-
5
-
8
-
RTS
With a hardware handshake:
▪
If connecting a thermostat to the PC, use a straight-through connection (1:1) and not
a null modem cable.
▪
The RS232 interface can be directly connected to the PC with a straight-through (1:1)
cable.
Without a hardware handshake:
▪
Set the corresponding operating mode on the PC.
▪
Use shielded connecting cables.
▪
Connect the shielding to the connector housing.
▪
Galvanically isolate the cables from the rest of the electronics.
▪
Do not connect up unused pins.