![Heidolph 506-21100-00 Скачать руководство пользователя страница 53](http://html1.mh-extra.com/html/heidolph/506-21100-00/506-21100-00_operating-instructions-manual_2118988053.webp)
54
Troubleshooting
Error codes on the device
Code
Description
E11
Sensor break, internal sensor
E12
Sensor break or short circuit, internal hardware
E13
Internal sensor difference < 15 °C
E14
Internal safety temperature (T
hotplate
> T
set
+ 25 °C)
E21
External temperature sensor: media contact lost in the
current process
E22
External temperature sensor without media contact when
switching on
E23
External safety temperature exceeded
E36
IO expander: Loss of communication
E41
Drive failure
E51
External sensor break
E52
External sensor difference > 15 °C
E53
External temperature sensor disconnected or plugged in
when the heating is switched on