Setup
Squareness screen
The Squareness screen contains data and choice fields
for calibrating the squareness of the measuring system.
The calibration of stage squareness requires the use of a
certified square artifact. To calibrate squareness:
1 Place the square calibration artifact on the stage
with one edge along the X-axis.
2 Measure the angle of the square calibration artifact
(ideally 90 degrees).
3 Select the Observed Angle data field, enter
the measured angle using the number keys and
press the Enter key to highlight the Standard
Angle field.
4 Enter the actual angle of the square calibra-
tion artifact obtained from the artifact’s certifica-
tion document into the Standard Angle data field.
5 Touch the Enabled choice field to toggle the
field to ON.
Encoders and Squareness Screens