BPro
®
BT (Connect to PC Version) Operator’s Manual
Starting BP session
19
8.4
Events to be mindful of during a BP session
1.
The BPro
®
BT will automatically trigger measurements at 15 minutes preset interval and give off
3 alarm beeps 3 seconds before starting each measurement.
2.
Whenever possible, keep the wrist still until the measurement is completed.
3.
The BPro
®
BT will display a tick to indicate a successful measurement.
4.
If the BPro
®
BT is unable to obtain a valid blood pressure reading, it will produce a long beep
and then activate a second measurement cycle. If the second attempt fails again, it will trigger
the third measurement cycle.
5.
The BPro
®
BT has been programmed to repeat this 3-cycle sequence two more times, with an
interval of thirty seconds between sequences, if no valid readings are obtained.
6. If the BPro
®
BT is unable to obtain valid blood pressure readings after a 3-cycle sequence.
Adjustment of sensor position is recommended.