Health O Meter 2650KL Скачать руководство пользователя страница 5

CERTIFICATIONS / CONNECTIVITY / DISPOSAL 

 

Certification Descriptions 

 

 

All Health o meter

®

 Professional scales are manufactured in FDA-approved factories. 

 

Scale Connectivity Information 

To enable reliable transmission of weight, height, and Body Mass Index (BMI) data, this scale is 
designed to connect to a computer, monitor, or other electronic device via a built-in USB port or 
Bluetooth wireless technology. This connectivity helps minimize a source of potential medical 
errors caused by incorrectly copying and then recording patient data.  For more information 
regarding how to connect the scale to other electronic devices, please contact Health o meter

®

 

Professional Scales Technical Support, available during normal business hours at 1-800-638-
3722. 

 

Scale Disposal 

 

 

 

Health o meter® Professional scales must be disposed of properly as electronic waste.  Follow 
the national, regional or local regulations which apply to you for disposal of electronic waste or 
batteries.  Do not dispose of this device in the domestic waste stream.  

 

 

 

 
 
 

 

Pelstar, LLC has been officially certified as the manufacturer 
of Health o meter

®

 Professional medical devices. The Pelstar, 

LLC quality assurance system covers the development, 
production, sales and service of medical scales and 
measuring systems. 

 

 

Complies with: 
ANSI/AAMI ES60601-1 (2005) + AMD 1 (2012), CAN/CSA-
C22.2 No. 60601-1 (2014), IEC 60601-1-6 (2010 ) + AMD 1 
(2013) 

E112697 

Содержание 2650KL

Страница 1: ...Wheelchair Scales User Instructions PELSTAR LLC 9500 West 55 th St McCook IL 60525 7110 USA www homscales com Pelstar LLC 2018 P N UM2650KL_2650KG_2700KL_2700KG_E_REV20180625 Equipped with Everlock E...

Страница 2: ...ounting Instructions 8 Powering the Scale 10 Set Up Instructions 11 Everlock Instructions 14 Operating Instructions 15 Transport Instructions 21 Maintenance and Cleaning 22 Calibration 23 Troubleshoot...

Страница 3: ...AC adapter cord for cracking fraying or for broken bent plug prongs Prior to using this scale ensure that the AC adapter is plugged into an outlet with the rated voltage appropriate for operation Ens...

Страница 4: ...0 lb x 0 2 lb 454 kg x 0 1 kg 2650KG 2700KG 454 kg x 0 1 kg Power Requirements 100 240V AC 50 60Hz to 9V DC Adapter Type UE15WCP 9 VDC 500 mA Item ADPT30 included or 6 D batteries not included Environ...

Страница 5: ...to other electronic devices please contact Health o meter Professional Scales Technical Support available during normal business hours at 1 800 638 3722 Scale Disposal Health o meter Professional scal...

Страница 6: ...ly remove each assembly from the carton and unwrap the packing materials Set the carton aside for storage To avoid damaging the scale parts when unpacking do not use a box cutter knife scissors or any...

Страница 7: ...level the scale Be certain that the feet are not completely tightened into the platform for proper load cell operation Open the scale to its operational position by unfolding the wheelchair ramps 3 C...

Страница 8: ...or placed on a table up to 71 1 8 m from the platform 1 Place the display head with its attached mounting bracket mount on the wall in the desired location Ensure that the display head is level and th...

Страница 9: ...1 8 3 2 mm from the wall Warning Extending the screw heads beyond 1 8 3 2 mm can damage the display assembly 6 Obtain the display head and the load cell cable Plug the load cell cable into the load c...

Страница 10: ...unscrewing the battery cover screw 2 Remove the battery pack from the display head Replace or install fresh batteries making sure to install them in the proper direction Health o meter Professional re...

Страница 11: ...is locked per the instructions above users can still view the alternate weight measuring unit by enabling the Unit View option Enabling the Unit View option allows the user to press the Unit button t...

Страница 12: ...time of inactivity before power off 1 Press to access the Help menu options 2 While in the Help menu press to select Set Options Press to advance to the next menu 3 While in the Set Options sub menu p...

Страница 13: ...n powered on rather it will use the last zero setting it had from its most recent use There will be no delay in using the scale if the Auto Zero is disabled Health o meter Professional recommends that...

Страница 14: ...1 800 815 6615 1 Press to access the Help menu options 2 While in the Help menu press to select Set Options Press to advance to the next menu 3 While in the Set Options sub menu press to scroll throu...

Страница 15: ...f a weight value to be deducted from the weight on the platform Also releases tare weight returns display to zero Toggles between pounds and kilograms If weight measuring unit is locked and Unit View...

Страница 16: ...ared weight of the object from the gross weight of the patient and object together The display will show the patient s weight 5 With no weight on the scale the display will show the tare value as a ne...

Страница 17: ...ent The display will show the patient s weight 5 The weight may be held on the display by pressing The scale will continue to display the patient s weight after they are off the scale and will remain...

Страница 18: ...on the display by pressing The scale will continue to display the patient s weight after they step off the scale and will remain on the display until released or the scale is powered off To release p...

Страница 19: ...patient remaining on the scale press 3 The scale will prompt entry of the patient s height Use the arrow keys to enter the patient s height Note When weighing in pounds the height is displayed in inch...

Страница 20: ...e parent s tared weight 4 Carefully place the child into the parent s arms The scale will automatically deduct the weight of the parent and only display the child s weight Note For an alternate method...

Страница 21: ...off of the mounting screws 3 Unplug the load cell cable from the scale platform and the display head Carefully store the load cell cable during transport 4 With the scale laying flat on the floor fold...

Страница 22: ...sinfecting Proper care and cleaning is essential to ensure a long life of accurate and effective weighing Caution Disconnect the scale from the AC adapter power source prior to cleaning the unit 1 Hea...

Страница 23: ...s should be used for calibration Warning Always ensure that the scale is returned to the correct UOM after scale calibration 1 Press to access the Help menu options 2 While in the Help menu press to s...

Страница 24: ...ibration weights to be placed on the platform Press after calibration weights are placed on the platform 9 The display will show Calibrating during the calibration process 10 If calibration is success...

Страница 25: ...balanced Completely hand tighten the feet then turn them back 1 full turns 4 Scale is out of calibration 4 Check weight with known weight value Weighing is performed but the weighing process takes too...

Страница 26: ...This warranty gives you specific legal rights and you may also have other rights that vary from country to country state to state province to province or jurisdiction to jurisdiction To get Warranty S...