User Manual
MK/YF38-07 – V2.01CS
© 2019 All Rights Reserved
Page 30 of 60
4.8 Treatment setting parameters
Parameter
Setting range
Description
Mode
Exit
Return to the previous menu
Patient Mode
Clinician Mode
Treatment
Mode
selection
CPAP
Auto CPAP
CPAP Mode: provides
continuous positive airway
pressure ventilation.
Auto-CPAP Mode (C5 only):
provides automatic adjustment
varying the positive airway
pressure as per the airway
obstruction state.
Clinician Mode
CPAP
4~20cmH20
Select the Treatment Pressure
based on the patient’s
prescription. This will be one
value between 4cmH20 and
20cmH20 in 0.5cmH20
increments.
Clinician Mode
Max CPAP
4~20cmH20
Set the upper pressure limit for
automatic CPAP ventilation
Clinician Mode
Min CPAP
4~20cmH20
Set the lower pressure limit for
automatic CPAP ventilation.
Clinician Mode
COMF
OFF, 1, 2, 3
Switch exhalation pressure relief
comfort technology OFF, or
choose a setting of 1, 2 and 3.
The higher the number, the
Patient Mode
Clinician Mode