56
TOP-Z30A/Z4A
A2212-1.0 de/en
HBM
Parameter
Value
1)
Unit
Change
requires
rights to
level:
Explanation
Excitation Level
(Minimum)
0.5
V
ID
Lower limit for the operating range
of the excitation voltage according
to the HBM data sheet.
Excitation Level
(Maximum)
12.0
V
ID
Upper limit for the operating range
of the excitation voltage according
to the HBM data sheet.
Calibration Date
1-Jun-2006
CAL
Date of the last calibration or
creation of the test certificate (if no
calibration carried out), or of the
storage of the TEDS data (if only
nominal values from the data sheet
were used). Format:
day-month-year.
Abbreviations for the months: Jan,
Feb, Mrz, Apr, Mai, Jun, Jul, Aug,
Sep, Okt, Nov, Dez.
Calibration Initials
HBM
CAL
Initials of the calibrator or calibration
laboratory concerned.
Calibration Period
(Days)
730
days
CAL
Time before recalibration, calculated
from the date specified under
Calibration Date.
Measurement location
ID
0
USR
Identification number for the
measuring point. Can be assigned
according to the application.
Possible values: a number from 0 to
2047. If that is not enough, the HBM
Channel Comment template can
also be used for this purpose.
1)
Typical values for a type TOP-Z30A/2kN force transducer from HBM
When the manufacturer creates the template
Bridge Sensor
, the physical
measured quantity and the physical unit are defined.
Unit available is specified in the IEEE standard for the respective measured
quantity. For example N for force.
At the time of creating the template it is also necessary to choose between the
options ”Full Precision”, ”mV/V” and ”uV/V” for the accuracy of the
characteristic transducer curve mapped in TEDS.
HBM always opts for ”Full Precision” here, in order to be able to use full digital
resolution. This choice is also recommended to users who program the TEDS
memory themselves.
For more information on TEDS please see the TEDS operating manuals
at our website www.hbm.com/TEDS