19
HBM: Business Document
6.1 Use of EMI suppressor
To suppress high frequencies a EMI suppressor on the cable between rotor
and stator has to be used. Use at least 3 loops of the cable.
Fastening must be done in an area not subject to mechanical loads (i.e. no
unwanted vibrations, etc.) using cable ties fit for the specific application.
Fig 6.1:
Installation example EMI suppressor
Information
Consider longer cable of approximately 40cm due to the installation of the EMI
suppressor. The max. distance of EMI suppressor to connector is 500mm.
If the EMI suppressor has to be removed for any purpose (e.g. for
maintenance), it must be replaced on the cable. Use only EMI suppressor of
the correct type.
Type: Vitroperm R
Model No.: T60006-22063W517
Size: external diameter x internal diameter x height = 63 x 50 x 25
Important
The use of the EMI suppressor on the power cable is mandatory to ensure
compliance with FCC regulations.