Mechanical installation
T40FH
A4429-3.0
HBM: public
23
Notice
Make sure during installation that you do not damage the measuring zone
marked in Fig. 5.5 by using it to support tools or knocking tools against it when
tightening screws, for example. This can damage the transducer and produce
measurement errors, or even destroy the transducer.
1. Prior to installation, clean the plane faces of the transducer flange and the
counter flange.
For safe torque transfer, the surfaces must be clean and free from grease.
Use a piece of cloth or paper soaked in solvent. When cleaning, make sure
that you do not damage the transmitter winding.
Hexagon screw DIN 933
Hexagon screw DIN 933
Measuring zone
Fig. 5.5
Bolted rotor connection
2. For connection of the flange, (
) use hexagon screws DIN 933 of
a suitable length (dependent on the connection geometry,
).
3. Fasten all screws with the specified torque (
).