![HBM T12 Скачать руководство пользователя страница 60](http://html.mh-extra.com/html/hbm/t12/t12_manual_2111944060.webp)
T12
60
A1979
−
10.0 en
HBM
Fieldbuses
CAN Bus
Protocol
−
CAN 2.0B, CAL/CANopen-compatible
Data rate
Meas.
values/
s
max. 4800 (PDO)
Hardware bus link
as per ISO 11898
Baud rate
kBit/s
1000
500
250
125
100
Maximum line length
m
25
100
250
500
600
Connector
−
5-pin, M12x1, A-coding per CANopen
DR
−
303
−
1 V1.3, electrically isolated from
power supply and measurement ground
PROFIBUS DP
Protocol
−
PROFIBUS DP Slave, per DIN 19245-3
Baud rate
MBaud
max. 12
PROFIBUS Ident Number
−
096C (hex)
Input data , max.
bytes
152
Output data, max.
bytes
40
Diagnostic data
bytes
18 (2
@
4-byte module diagnosis)
Connector
−
5-pin, M12x1, B-coding, electrically
isolated from power supply and
measurement ground
Update rate
6)
Meas.
values/
s
Configuration entries
v
2
4800
v
4
2400
v
8
1200
v
12
600
v
16
300
u
16
150
Limit value switches (on fieldbuses only)
Number
−
4 for torque, 4 for rotational speed
Reference level
−
Torque low pass 1 or low pass 2
Rotational speed low pass1 or low pass 2
Hysteresis
%
0 to 100
Adjustment accuracy
digits
1
Response time (LP1 = 4000 Hz)
ms
typ. 3
TEDS (Transducer Electronic Data Sheet)
Number
−
2
TEDS 1 (torque)
−
A choice of voltage sensor or frequency
sensor
TEDS 2 (rotational speed/angle of
rotation)
−
Frequency/pulse sensor
6)
When CAN PDOs are activated simultaneously, the update rate on the PROFIBUS is reduced.