T12
50
A1979
−
10.0 en
HBM
Template: High Level Voltage Sensor
Parameter
Value
Unit
Required
user
rights
Explanation
Minimum Torque
0.000
N
@
m
CAL
The physical measurand and
unit are defined when the
template is created, after which
they cannot be changed.
Maximum Torque
1000
N
@
m
CAL
Minimum Electrical Value
0
V
CAL
The difference between these
values is the nominal (rated)
sensitivity.
Maximum Electrical Value
10
V
CAL
Discrete Signal Type
Bipolar
ID
Discrete Signal Amplitude
5
V
Discrete Signal
Single
Transducer Response
Time
0
Excitation Level nom
24
V
Excitation Level min
18
V
Excitation Level max
30
V
Excitation Type
DC
Excitation Current draw
0.5
A
Calibration Date
1-Nov-2006
CAL
Date of the last calibration or
creation of the manufacturing
certificate (if no calibration
carried out), or of the storage of
the TEDS data (if only nominal
(rated) values from the data
sheet were used).
Format: day-month-year.
Abbreviations for the months:
Jan, Feb, Mar, Apr, May, Jun,
Jul, Aug, Sep, Oct, Nov, Dec.
Calibration Initials
HBM or
PTB
CAL
Initials of the calibrator or
calibration laboratory
concerned.
Calibration Period (Days)
0
days
CAL
Time before recalibration,
calculated from the date
specified under Calibration
Date.
Measurement Location ID
0
USR
Identification number for the
measuring point. Can be
assigned according to the
application. Possible values: a
number from 0 to 2047.