Connections
Observe the safety instructions before commissioning the device.
2.1 Factory settings
Before operating the device, check the parameters set at the factory and note
that the elements for selecting the analogue output signal (current/voltage
output) and for setting synchronisation, are located on the motherboard.
The factory settings are given below:
D
Mains voltage: 230 V / 50...60 Hz or 115 V / 50..60 Hz, depending on order
D
Analogue output: output voltage
"
10 V
D
Synchronisation: master
ST14 ST15
Analogue output:
Voltage
Current
ST13
ST13 ST14 ST15
ST100: for attaching spare bridges
ST9 and ST10:
for options
ST11
ST11
Master:
Slave:
Master/Slave setting:
ST11
ST10
ST13
ST9
IC
IC
ST15
ST14
Transformer
Frontsite
Fuses (slow-blowing)
ST100
Fig. 2.1:
Location of jumpers on motherboard
2.2 Changing the factory settings
To change the factory settings, proceed as follows:
1 Switch off the device and take out the mains cable. Remove all the plug
connections on the back panel.
2 Loosen the four screws on the cover of the housing and remove the cover.
3 Change whichever setting is relevant to you with the aid of the jumpers, by
following Fig. 2.1
4 Screw the cover of the housing back in position.