12
PW20i
HBM
A1269-2.0 en/de/fr
The current measured value is read out using the command MSV?. The format
of the measured value (ASCII or binary) is set with the command COF. You
can also use command COF to select automatic data output.
Different types of digital filter are implemented in the PW20i and these are
selected using the FMD command. With FMD0, filters are also available below
the 1 Hz cut-off frequency. In filter mode FMD1, fast-settling filters with high
damping in the stop band are activated. Detailed information can be found in
the help file (”Individual command descriptions”).
5.3
Trigger function
The PW20i includes four trigger functions to support measurements in
packaging machinery and checkweighers:
•
Internal pre-triggering via an adjustable level
•
External triggering via the digital control input TRG (pre-trigger)
•
Internal post-triggering via an adjustable level
•
External triggering via the digital control input TRG (post-trigger)
Gross or net values can be processed as measured values (TAS).
The trigger function is described in the help file, Trigger function overview.