Electrical connection
12
A03392_02_Y00_00
HBM: public
PW15PH
5
Electrical connection
The following can be connected for measurement signal conditioning:
S
Carrier-frequency amplifiers
S
DC voltage amplifiers
that are designed for strain gage measurement systems.
The load cells are supplied in the standard version with an 8-pin connector in a
six-wire configuration. The load cells are optionally available with different
cable lengths and with a hygienic design connection cable.
5.1
Connection in a six-wire configuration
Plug-in contact 1 = measurement signal (+)
Plug-in contact 6 = excitation voltage (-)
Plug-in contact 5 = sense lead (-)
Plug-in contact 7 = excitation voltage (+)
Plug-in contact 3 = sense lead (+)
Plug-in contact 8 = measurement signal (-)
2
3
4
5
6
7
1
8
Plug-in contact 2 = not in use
Plug-in contact 4 = not in use
Fig. 5.1
Connector pin assignment
With this connector pin assignment, the output voltage at the amplifier is
positive when the transducer is under load (
). You can find the pin
assignment for cables available as accessories in
Section 9 on page 18
Содержание PW15PH Series
Страница 1: ...Mounting Instructions Montageanleitung Notice de montage English Deutsch Fran ais PW15PH...
Страница 3: ...Mounting Instructions Montageanleitung Notice de montage English Deutsch Fran ais PW15PH...
Страница 22: ...Accessories 20 A03392_02_Y00_00 HBM public PW15PH...
Страница 23: ...Mounting Instructions Montageanleitung Notice de montage English Deutsch Fran ais PW15PH...
Страница 43: ...Mounting Instructions Montageanleitung Notice de montage English Deutsch Fran ais PW15PH...
Страница 63: ...Accessoires PW15PH A03392_02_Y00_00 HBM public 21...