PME-MP60/MP07
A0616-13.4 en
HBM
Group
Parameters
Meaning
ANALOG
OUTPUT
SourceVo
You can choose the gross or net values, as well as the peak
value as the source of the analogue signal.
Mode Vo
Use DIP switches S11 to specify the signal mode for the
analogue output. The following options are available:
"
10 V,
"
20 mA, 4 - 20 mA
Zero %
1)
Zero V
EndV
kNm
1)
EndV V
Physical unit
V
0
Zero kNm
Zero V
EndV kNm
EndV V
Information on scaling
Output characteristics:
The scale factor for the analogue output is a result of the
input and output characteristics. If the set nominal value
corresponds to the measuring range in kHz, then the
minimum output voltage to be set is 0.5 V. In the case of
settings which lead to exceeding the respective limits, you are
given the message ”Analogue scaling error” (see page 68).
Scale range analogue output min.: 0.5 V at 100 % of input measuring
range
Scale range analogue output max.: 10 V at 1 % of input measuring
range
1)
Depending on the unit selected