![HBM MP07 Скачать руководство пользователя страница 10](http://html1.mh-extra.com/html/hbm/mp07/mp07_operating-manual_2111883010.webp)
10
PME-MP60/MP07
A0616-13.4 en
HBM
Note:
You must set amplifier type MP60 on DIP switch S12 (MP60)
(see also page 13).
1 2 3 45 6
S12
ON
Switch positions
must not be
changed!
Module MP07:
The MP07 module is used for the excitation of torque transducers.
The following HBM torque transducers may derive their excitation from an
MP07 module:
T10F-KF1, T30FNA, T32FNA, T34FN, T36FN
The MP07 module, like the MP60, is connected to a 24 V supply voltage and
delivers the following output voltages (short-circuit proof):
+15 VDC, 100 mA
-15 VDC, 100 mA
54 V
PP
/ 75 V
PP
, 150 mA (24 - 25 kHz square-wave excitation voltage)
The supply voltage is electrically isolated from the calibration signal inputs.
The calibration signal is triggered by increasing the excitation voltage of the
transducer shaft from 54 V
PP
to approx. 75 V
PP
1)
.
The MP07 module has no display and no control keys.
1)
See operating manual for torque transducer