PME-MP60/MP07
A0616-13.4 en
HBM
Enter T10F characteristic values:
Group: TRANSDUCER
Unit:
U/min
Input range:
1 MHz
F2
ON
Zero index (+)
OFF
Frq x 4
ON
1)
Direction of rotation
ON
(only possible if F1 and F2 are active)
Switching threshold
2.5 V
Glitch filter
ON
Zero kHz
0
Zero U/min
0
NVal kHz
144 kH
z
NVal U/min
6000 U/min
Trans Err
OFF
1)
every slope is counted, not just the pulses; this quadruples the resolution
NOTE
For frequency quadrupling, you have to adapt the input characteristics.
144 kHz (instead of 72 kHz with F2 = OFF) then correspond to a
revolution count of 6000 U/min.
Switch:
F2
OFF
ON
OFF
ON
Switch:
Frq x 4
OFF
OFF
ON
ON
Frequency display
1x
1x
2x
4x
Detection of direction
of rotation
no
yes
no
yes