Remote Communications
60
A4249-8.0
HBM: public
FS22 SI
:ACQUisition:CONFiguration:RANGe:FOR
Mula:C:NR:[CWL;FML],...,[CWL;FML]
Defines the formula(s) and central wavelength(s) for the
range(s) on channel C. NR defines the number of ranges in
channel C, CWL represents the central wavelength of the
sensor in the specific range and FML represents the corre
sponding formula.
Note: This command shall be used after the ranges have
been defined with the :ACQUisition:CONFigura
tion:RANGe:WAVElength:C:NR:MIN,MAX,...,MIN,MAX com
mand.
Note: The central wavelengths and formulas shall be sorted
by wavelength in ascending order.
Note: The variable "x" is used in the formulas to represent
the wavelength variation of the sensor relatively to the cen
tral wavelength.
Example:
Command
:ACQU:CONF:RANG:FORM:1:2:[1512.1
;-96.2*x^2+104.8*x+30],[1520.5;x*0.78]
Answer
:ACK
:ACQUisition:CONFiguration:RANGe:FORMula:C:RN:FML
Defines the formula for the range number RN on the chan
nel C. Range Number RN is an integer from 1 to the total
number of ranges NR.
Example:
Command
:ACQU:CONF:RANG:FORM:3:2:-13.2*x
^2+104.8*x+30
Answer
:ACK
Содержание FS22 SI
Страница 1: ...User Manual English FS22 SI Industrial BraggMETER SI...
Страница 22: ...Operation 22 A4249 8 0 HBM public FS22 SI Fig 3 6...
Страница 23: ...Operation FS22 SI A4249 8 0 HBM public 23 Set the IP address and the subnet mask as in Fig 3 7 Fig 3 7 Press OK...
Страница 92: ...BraggMONITOR SI User Interface 92 A4249 8 0 HBM public FS22 SI Fig 7 17 Fig 7 18...
Страница 93: ...BraggMONITOR SI User Interface FS22 SI A4249 8 0 HBM public 93 Fig 7 19 Fig 7 20...