Command set of the HBM interpreter
Amplifier functions
58
A03474_04_E00_00
HBM: public
DMP41
Information
This command applies to all channels of a device.
Information
The measured values are scaled in ASCII format or output in ADU units in
binary format.
Change Output Format Query
Query measured value output format
Syntax:
COF?(x)
Parameter:
none
Effect:
Code number of measured value output format is output.
Response:
q1(y)
Code number of the output format (see COF command).
Example:
COF?(x)
0(y)
Measured values, Channel No. and Status are output in
ASCII format.
Input Sampling Rate
Set measured value transfer rate
Syntax:
ISR p1(x)
Parameter:
p1
Divider value for the data rate with measured value output
via an interface.
Effect:
This command ensures output in a fixed timeslot for mea
sured value output using the command MSV?x,y.
The relationship between parameter p1 and the output rate
results from the calculation below.
The following calculation applies for the determination of the
input value for p1:
p1 =
Internal cycle frequency
Measured value transfer rate
(=75 Hz)
=
1...75
COF?
ISR
Содержание DMP41
Страница 1: ...Interface description and command set English DMP41 Digital precicion measuring device...
Страница 10: ...Interface comparison 10 A03474_04_E00_00 HBM public DMP41...
Страница 14: ...Communication with the DMP Ethernet interface 14 A03474_04_E00_00 HBM public DMP41...
Страница 18: ...Communication with the DMP RS 232 interface 18 A03474_04_E00_00 HBM public DMP41...
Страница 22: ...Program creation Communication via the Ethernet interface Test program 22 A03474_04_E00_00 HBM public DMP41...
Страница 86: ...Functional command overview 86 A03474_04_E00_00 HBM public DMP41...
Страница 90: ...Index 90 A03474_04_E00_00 HBM public DMP41...
Страница 91: ...Index DMP41 A03474_04_E00_00 HBM public 91...