![HBM DMP41 Скачать руководство пользователя страница 44](http://html1.mh-extra.com/html/hbm/dmp41/dmp41_interface-description_2111867044.webp)
Command set of the HBM interpreter
Amplifier settings
44
A03474_04_E00_00
HBM: public
DMP41
Acknowledge-
ment
Significance
0
Command executed
?
Error
Example:
Input of cut‐off frequency and filter characteristic for filter
fc2.
ASF 2,4,0 (x)
0(y)
Filter fc2 is set to 0.22Hz cut‐off frequency and
Bessel characteristic.(see tables for command ASF?)
Amplifier Signal Filtering Query
Output of cut‐off frequency and filter characteristic
Syntax:
ASF? p1(x)
Parameter:
p1
Filter code number
0
Frequency table (Bessel and Butterworth)
1
Filter fc1
2
Filter fc2
Effect:
Output of cut‐off frequency and filter characteristics set in
the low‐pass filter.
Response:
If p1=0
q1, q2(y)
q1
Table, Bessel frequencies
q2
Table, Butterworth frequencies
If p1
≠
0
q1, q2, q3(y)
q1
Filter number fc1 / fc2
q2
Cut-off frequency of filter fc1/fc2
q3
Filter characteristic (0=Bessel, 1= Butterworth)
Example 1:
Table of possible filter frequencies
ASF?0(x)
“
0.040 0.080 0.100 0.200 0.0400 0.800 1.000
”
,
“
2.000 4.000 8.000 10.000 20.000 40.000
”
(y)
ASF?
Содержание DMP41
Страница 1: ...Interface description and command set English DMP41 Digital precicion measuring device...
Страница 10: ...Interface comparison 10 A03474_04_E00_00 HBM public DMP41...
Страница 14: ...Communication with the DMP Ethernet interface 14 A03474_04_E00_00 HBM public DMP41...
Страница 18: ...Communication with the DMP RS 232 interface 18 A03474_04_E00_00 HBM public DMP41...
Страница 22: ...Program creation Communication via the Ethernet interface Test program 22 A03474_04_E00_00 HBM public DMP41...
Страница 86: ...Functional command overview 86 A03474_04_E00_00 HBM public DMP41...
Страница 90: ...Index 90 A03474_04_E00_00 HBM public DMP41...
Страница 91: ...Index DMP41 A03474_04_E00_00 HBM public 91...