9
Z6/M
MOUNTING
Notice
If the load cell is to be replaced, the upper module plate can be lifted by max. 1.5 mm.
Lock the nut in place to hold the upper module plate in this position. Dismounting the tank
itself is not required.
The following tightening moments prescribed for the fastening bolts must be adhered to
when replacing the load cell:
up to 200 kg max. Capacity
Galvanized version:
34 N·m
Stainless steel version:
23 N·m
500 kg max. Capacity
Galvanized version:
76 N·m
Stainless steel version:
45 N·m
The setting of the overload stop must be checked after the replacement load cell has been
installed. Please refer to the specifications for the weighing module: ”Setting range for the
overload stop”.