AD103C
ELECTRICAL CONNECTION
14
6
ELECTRICAL CONNECTION
6.1
Transducer connection
SG transducers with full bridge can be connected.
Bridge resistance R
B
= 40...4000
Ω
(external excitation voltage). SG transducers must
operate at a bridge excitation voltage of 5 V
DC
.
Transducers with a bridge resistance >1000
Ω
can also be connected in principle. How
ever, this will increase the noise of the measurement signal (increased measurement
ripple).
Currentconsumption
+
≤90mA
)
ExcitationvoltageU
B
BridgeresistanceR
B
Important
Make sure that a low-noise constant voltage source is used for the excitation voltage, as
the quality of the power supply is directly reflected in the measurement result (see Calcula
tion).
6.1.1
Connection in a 6-wire configuration
HBK color
code
Pin
socket
Bridge excitation voltage (+)
bl
9/21
Sense lead (+)
gn
10
Measurement signal (+)
wh
24
Measurement signal (–)
rd
11
Bridge excitation voltage (–)
bk
12/13
Sense lead (–)
gy
23
Cable shield
ye
Housing
Fig. 6.1
Transducer connection (6-wire) to the AD103C amplifier board
14
1
gy
wh
25
bl
gn
rd
bk
13
Bu1 amplifier
(back of connector)