No.:GC-JS-21 Version: 01/00
Effective date
:
2020-03-30
39
The time
difference is
Total Cycle time
(18 m : 34 s)
SERIAL
NO.:300012345
OPERATER:
----CLASS B---
COUNT:00001
B.W.
WATER
COUNT
DATE:
01-07-
2020
TIME: 15:46:25
PROGRAM: 134 /
QUICK
HH:MM:SS
KPA ℃
Temperature
Unit
START
15:51:43 000 072.6 05:18
VACUUM
15:52:45 -080 079.1 01: 02
Time of vacuum phase
RAISE
15:53:00 021 089.2 00:15
VACUUM
15:53:50 -080 095.2 00:50
RAISE
15:54:05 010 103.9 00:14
VACUUM
15:55:09 -080 105.7 01:05
RAISE
15:56:24 221 133.7 01:15 STER-START:
15:56:24 221 133.7 01:15
MAX TEMP:
136.1
MIN TEMP:
133.7
During sterilization
MAX PRESS:
227
phase
MIN PRESS:
207
STER-END
16:00:24 213 134.6 04:00
DRY-END
16:04:58 -017 118.8 04:34
CYCLE COMPLETE
16:04:59 000 119.1 00:01
Appendix 3
:
Explanation of each stage
No.
Display
Explanation
1
Standby
The device is normal and ready for operation
2
Warm up
Warm-up phase
3
vacuum
Vacuuming
4
Temperature rise
Steam is being injected into the sterilization chamber: temperature and
pressure will increase
5
drain
The sterilization room is draining
6
Sterilization
Sterilizing: maintain sterilization temperature and pressure
Pressure Unit