7.6 Choose a program
Choose the appropriate sterilization program according to the type of load. For the sterilization load type
corresponding to each program, see chapter 5.2.1
There is no difference in sterilization effect between the 134 ° C sterilization procedure and the 121 ° C
sterilization procedure. Try to use the 134 ° C procedure to shorten the sterilization time.
Solid program can only sterilize solid instruments that are not wrapped, and should be used immediately after
sterilization.
This sterilizer is only suitable for the sterilization of moisture- and heat-resistant medical devices, not
for the sterilization of oils and powders such as petroleum jelly and agar.
This sterilizer cannot be used for sterilization of liquids and liquids sealed with closed containers
(especially glassware), which may cause the containers to burst, endangering the safety of persons and equipment.
7.7 Run the program
After the user selects the program in the user menu or directly selects the program by pressing the shortcut bar on
the home page, the interface to be run will come out
:
Press "Start" to run the program automatically.
7.8 Man-made interruption
If the user needs the device to stop running during the operation of the device, a confirmation dialog box will appear
after pressing the "Stop" key on the running interface
: