No.:GC-JS-21 Version: 01/00
Effective date
:
2020-03-30
22
7
Wait machine to
cool down then
try again
The sterilization chamber was too hot when the 134 ° C
program was just run, and then the 121 ° C program was run.
The door needs to be opened for cooling. This time takes up
to 5-7 minutes
8
Please
replace
bacterial filter
The life of the bacteria filter has expired and needs to be
replaced
9
Please
replace
door seal
The door seal has reached the end of its life and needs to be
replaced
10
Need
maintenance
The periodic maintenance time has expired, and the
manufacturer or manufacturer's representative is required to
maintain the equipment
11
Last cycel
ended
improperly
Since the last sterilization cycle was stopped by the user, the
power failure or failure did not end normally, the load needs
to be re-sterilized
12
Please clear error
Equipment failure equipment cannot be used. Qualified
personnel are required to confirm whether the equipment is
intact. If there is a problem, it needs to be repaired before it
can be used after clearing the fault.