VPT10-H
HART
®
PRESSURE TRANSMITTER
INSTALLATION, OPERATION, CONFIGURATION AND MAINTENANCE MANUAL
______________________________________________________________________________________________________________________________
10
2
INSTALLATION
RECOMMENDATION
When taking the equipment to the installation location, transfer it in the original packaging. Unpack the
equipment at the installation location to avoid damage during transportation.
RECOMMENDATION
Model and specification of equipment are indicated on identification plate, located at the top of the
housing. Check if supplied specification and model correspond to application requirements. Be aware of
the maximum and minimum specifications and sensor range. After installation in the field, see Calibration
topic.
STORAGE
The following precautions should be observed when storing the equipment, especially for a long period:
(1) Select a storage area that meets the following conditions:
a) No direct exposition to rain, water, snow or sunlight.
b) No exposition to vibration and shocks.
c) Normal temperature and humidity (around 20°C / 70°F, 65% RH).
However, it can also be stored under the following temperature and humidity intervals:
●
Ambient Temperature: -40°C to 85°C (without LCD)* or -30°C to 80°C (with LCD)
●
Relative Humidity: 5% to 98% RH (@ 40°C)
(2) For equipment storage, use original factory package (or similar).
(3) If storing an already used Vivace equipment, dry every moist part and clean all connections that
was in contact with the process. Keep covers and connections closed and properly protected for its
specific application and requirements.
* Only for general use. For explosion proof version, follow product certification requirements.