HARRIS PRODUCTS GROUP

  

920 SERIES

GAS DISTRIBUTION EQUIPMENT

 

Installation and Operation Manual

 

SWITCHOVER SYSTEM

   

THIS BOOKLET CONTAINS PROPRIETARY INFORMATION FROM THE HARRIS PRODUCTS GROUP.  IT IS PROVIDED TO THE PURCHASER SOLELY FOR 
USE IN CONJUNCTION WITH AUTOMATIC SWITCHOVER MANIFOLDS MODEL 920 SERIES. 

General Precautions – 

•  Installation should be performed by personnel knowledgeable in the handling of compressed gases.
•   Choose a location well ventilated for installation 
•   Use proper precautions when dealing with any compressed gas.
•   The user should be familiar with the chemical and reactive properties of the gas being used (Read the MSDS of the gas being used).

Reference Material – 

The following safety bulletins are available from the Compressed Gas Association (

www.cganet.com

•   SB-2 Oxygen Deficient Atmospheres
•   SB-10 Correct Labeling & Proper Fittings On Cylinders/Containers
•   SB-15 Avoiding Hazards In Confined Work Spaces During Maintenance, Construction And Similar Activities
•   SB-19 Potential Valve Thread & Cylinder Thread Mismatch

General Instructions - 

Switchover Manifolds should be installed in accordance with guidelines stated by the National Fire Protection Association, the Compressed Gas 
Association, OSHA and all applicable local codes. Switchover manifolds should not be placed in a location where the temperature will exceed 
120°F (49°C) or fall below –20°F (-29°C) for sustained periods of time.  A manifold placed outside in an open location should be protected against 
rain and excessive moisture.  During winter, protect the manifold from ice and snow.  The manifold should be located in a clean, well-ventilated 
area which is free of oil and combustible materials.

Introduction

The changeover panel is designed to provide continuous uninterrupted gas flow for applications using cylinder gas supplies up to 3000 PSIG. Ideal 
for high purity applications, this panel consists of two high purity primary and line regulators, gauges, isolation valves, and mounting panel.

Unscheduled downtime created by depleted cylinders can be eliminated with this panel incorporated into your system. When the primary side is 
consumed, the panel automatically switches to the reserve supply and feeds the process at a constant pressure. The line regulator eliminates 
virtually any pressure variation associated with this switch.

Maximum Inlet Pressure:

 

3000 PSIG 

Outlet Pressure:  

10 - 100 PSIG or 20-200 PSIG (User Adjustable)

Maximum Flow Rate:

  

200 SCFH

Содержание 920 Series

Страница 1: ...h guidelines stated by the National Fire Protection Association the Compressed Gas Association OSHA and all applicable local codes Switchover manifolds should not be placed in a location where the tem...

Страница 2: ...o not allow any oil grease or other potential contaminating lubricants to come in contact with any components These panels are factory assembled under rigorous high purity conditions without any use o...

Страница 3: ...below the set point of the other regulator The other side will then be operating at a higher pressure which will force the system to switch over and supply gas to the downstream process Model 920 Spec...

Страница 4: ...is properly cleaned to prevent the introduction of contaminants into the gas stream Harris Equipment Warranty Policy Harris warrants that manifold products sold in non corrosive gas service are free f...

Страница 5: ...ablecidas por la Asociaci n Nacional de Protecci n contra el Fuego la Compressed Gas Association OSHA y todos los c digos locales aplicables Los colectores de conmutaci n no deben colocarse en un luga...

Страница 6: ...xi n CGA de ajuste manual opcional No permita que ning n aceite grasa u otro lubricante contaminante entre en contacto con ning n componente Estos paneles se ensamblan en f brica en condiciones riguro...

Страница 7: ...ebajo del punto de ajuste del otro regulador El otro lado funcionar a una presi n m s alta lo que forzar al sistema a conmutar y suministrar gas al proceso posterior Especificaciones del modelo 920 Pr...

Страница 8: ...de gas Pol tica de garant a de equipos Harris Harris garantiza que los productos de colector vendidos para el servicio de gas no corrosivo est n libres de defectos de fabricaci n en materiales y mano...

Страница 9: ...a valve et le filetage du cylindre Consignes g n rales Les collecteurs commutateur doivent tre install s conform ment aux recommandations de la National Fire Protection Association la Compressed Gas A...

Страница 10: ...l ou la main en option Ne pas laisser d huile de graisse ou d autres lubrifiants potentiellement contaminants entrer en contact avec les composants Ces panneaux sont assembl s en usine dans des condit...

Страница 11: ...ndeur L autre c t fonctionnera alors une pression plus lev e ce qui forcera le syst me commuter et fournir du gaz au proc d en aval Sp cifications pour le mod le 920 Pression d admission nominale maxi...

Страница 12: ...la p n tration de contaminants dans le flux de gaz Politique de garantie de l quipement Harris Harris garantit que les collecteurs vendus pour le service au gaz non corrosif sont exempts de d fauts d...

Отзывы: