![Hardy Instruments Any-Weigh Скачать руководство пользователя страница 27](http://html.mh-extra.com/html/hardy-instruments/any-weigh/any-weigh_operation-and-installation-manual_597397027.webp)
Any-Weigh™ Series Floor Scales
19
FIG. 14 INSERTING THE LEVELING FOOT
INTO THE LOAD SENSOR
NOTE:
You may need to push with a little force due to some
interference caused by the O-ring that is attached to
the foot. DO NOT USE A HAMMER OR TRY TO
DRIVE THE FOOT INTO THE HOUSING.
Step 3.
With the through holes aligned, insert the
lock bolt going from the center of the scale
outward. This eliminates having to remove
the load sensors when installing the lock
bolt.
Step 4.
Screw the nut onto the lock bolt until it is
tight. The feet should look like the follow-
ing: (See Figs. 15 & 16)