Setting Fill Levels
The calibration for fill on this unit is intended to be set for a standard 1”x3”x1mm glass slide. We
recommend using ProSlide™ Frosted Microscope Slides (
PF72P
).The fill levels can be
customized to fit the operator’s preferences. We recommend a fill level reset after each new
tubing kit installation. The following steps will guide you through recalibrating the fill level on
your HemaPRO unit after a tubing change.
You may also view our video available on Hardy Diagnostics’s YouTube page for additional
assistance:
HemaPRO Set Fill Levels -
https://www.youtube.com/watch?v=90IUWF10Jhs
.
Fill levels set without a 1”x3”x2mm slide inserted in the cuvette during programming
will result in incorrect fill levels. This will lead to overflow of the reagents onto the
operating area. This can lead to possible damage of the unit’s surroundings.
Ensure the unit has been primed twice, as explained in
Tubing Kit
Replacement
(step 14) before resetting fill levels.
Power the instrument off.
Place a blank 1”x3”x1mm slide into the cuvette. A slide must be loaded before
fill levels can be adjusted.
While the machine is off, hold down the
<F4>
button.
While still holding the
<F4>
button, power the instrument back on and release
the
<F4>
button.
Once the instrument is powered back on, you will see the above service
screen. Select the
<F4>Yes
button to proceed.
NOTE
: Selecting
<F1>No
will return you to the Main Menu.
Set Fill Volume?
F1=No F4=Yes