IFU-10891[C] HemaPRO
User
Manual
Page 21 of 33
12.1.3.1 Place the “
Stain
” line that has a
blue tag
into the Wright-Giemsa Stain
bottle.
12.1.3.2 Place the “
Buffer
” line that has a
green tag
into the Wright-Giemsa Buffer
bottle.
12.1.3.3 Place the “
Rinse
” line that has a
white tag
into the Wright-Giemsa Rinse
bottle.
Remove the Stain Activation Module (the short cord that resembles a phone
cord) from the HemaPRO reagent kit box. Connect the Stain Activation Module
to the receptacle on the right-hand side of the instrument.
NOTE
:
If the Stain Activation Module is not plugged in when the unit is turned on,
the below message will display, accompanied by an audio alert. Turn unit off,
connect Stain Activation module and turn unit on to continue.
Power the instrument back on. The instrument will continue with its normal boot
cycle, and will land upon the Main Menu, below:
Main
Menu
Before using the instrument, you will need to
<F1> Prime
the unit twice.
NOTE
:
For every new reagent kit and tubing kit that has been installed on the
HemaPRO, a double Prime will be required. This is to ensure that all reagent
lines are properly filled and ready for staining.
Stain Activation Module
The HemaPRO is programmed to display two different warning screens related to the
Stain Activation Module. Please review the descriptions below to understand how each screen
differs and how to address these warnings.
Reagent Pack Empty
Self Test Failure:
Pack not connected
Cycles Remaining 125
F1=Prime F4=Cont.
Reagent Pack Empty
change pack F4=Cont.