Tubing Kit Replacement
The orange pump tubes and the clear lines with color coded cannulas must be replaced
every six (6) months to ensure the HemaPRO is operating under optimal conditions. The
following steps will guide you through replacing the tubing on your HemaPRO unit.
You may also view our video available on Hardy Diagnostics’s YouTube page for additional
assistance:
HemaPRO Tube Replacement -
https://www.youtube.com/watch?v=ntmFMWqefqI
To begin this procedure:
Empty the waste container. Be sure the line labeled “
Waste
” remains in the
waste container after it has been emptied.
Remove the three reagent cannulas labeled “
Stain
”, “
Buffer
”, and “
Rinse
” from
their respective bottles and wipe each cannula with an isopropyl alcohol wipe.
Hold them all at the opening of the empty waste container.
NOTE:
The three reagent cannulas should not be placed in the waste
container, as they may pull back in the fluid that was just expelled from
the waste line.
From the Main Menu:
Proceed to run two (2) prime cycles by pressing
<F1>Prime
while these
cannulas are suspended above the waste container. The purpose of this step is
to purge all remaining fluid from the tubing kit.
With all tubes now empty, power the instrument off.