IFU-10891[C] HemaPRO
User
Manual
Page 15 of 33
The unit will now prompt you to prepare the cuvette before inserting the slide.
Continue to prepare the cuvette by pressing
<F4>
.
NOTE
: “Preparing” will fill the cuvette about half way with stain. Once filled, the
below screen will show.
When prompted, you may now load the slide into the cuvette with the specimen
facing out towards the front (facing you). Once the slide has been placed inside
the stain filled cuvette, immediately press the
<Stain>
. The unit will proceed to
stain the slide, followed by the buffer, and quick rinse of the slide.
When the staining process is complete, promptly remove the slide from the
cuvette and press the
<F4> Cont.
to continue to the Main Menu.
NOTE
:
DO NOT
leave the slide in the cuvette after staining. The stain quality is
best when the slide is promptly removed from the cuvette and allowed to air dry.
Main
Menu
NOTE
: When the unit returns to the Main Menu, your cycle count will be reduced
by 1.
To Prepare Cuvette:
(F1=back) Press F4
WAIT !!!
Preparing Cuvette
NOW load slides THEN
press STAIN to Time!
Done: Remove Slides
F4=Cont.
Cycles Remaining 124
F1=Prime F4=Cont.